Abstract
Background: An MRI is potentially hazardous for patients with retained ferromagnetic bullets. Recent studies have aimed to develop dual-energy computed tomography (DECT) as a screening tool for recognising highly ferromagnetic bullets. Inconsistent findings have been ascribed to inherent CT technology differences. Previous research demonstrated significant Hounsfield unit (HU) measurement variation among single-source CT machines.
Objectives: This study investigated the theoretical dual-energy index (DEI) variation between DECT machines when evaluating the potential ferromagnetic properties within the same sample of ex vivo bullets and metal phantoms.
Method: An experimental ex vivo study was conducted on eight metal phantoms and 10 unused bullets individually positioned in the same Perspex head phantom and scanned on two DECT machines. Two senior radiology registrars independently recorded the HU readings, and DEI values were calculated. Statistical analysis was performed using non-parametric methods for paired data, namely the Signed Rank Test. The DEI values based on mean HU readings between the DECT machines were compared.
Results: Inter- and intra-reader agreement was not statistically significant. The metal phantoms had poor interscanner agreement, with an overlap of the ferromagnetic and non-ferromagnetic ranges. The bullets had good interscanner agreement, with a similar ferromagnetic to non-ferromagnetic relationship.
Conclusion: The use of DEI values negates the previous assumption that significant interscanner variability exists among different DECT technologies while assessing highly attenuative ex vivo bullets.
Contribution: This investigation demonstrated that even though HU readings may be variable, the implementation of the DEI equation translates this into comparable values with good interscanner agreement.
Keywords: dual-energy computed tomography; ferromagnetic bullets; DECT machines; inter- and intra-reader agreement; MRI safety.
Introduction
South Africa has a very high violent crime burden, related to high morbidity and mortality rates. According to the South African crime statistics released for 2019–2020, firearm violence resulted in 7351 murders (of a total 21 325) and 12 718 attempted murder cases (of a total 18 635). Firearms were further implicated as the main tool of intimidation in most other crime categories and established relations with multiple murders.1
Law enforcement and criminal interactions are not the only source of gunshot injuries, as these may also occur during sport and hunting. Non-fatal gunshot injuries are widespread, and as a result, penetrating trauma makes up a large portion of the trauma population requiring medical care.2,3 Recent research highlighted gunshot wounds as a leading cause of spinal cord injuries, mostly involving younger males. Patients experienced a higher rate of adverse medical complications and morbidity, often required surgery, and had prolonged hospital admissions.4
Continuous advances and improvements in emergency medical care and technology will likely decrease trauma fatalities. It can reasonably be concluded that secondary evaluation techniques and residual lesion identification will become increasingly important as patients require further and definitive care.5
An MRI was previously deemed an absolute contraindication for patients with retained bullets, especially if these were close to vital structures. The main concerns are the potential hazards of in vivo position shift and the heating effect of the bullet within the MRI magnetic field. A recent investigation found that ferromagnetic MRI-related incidents are underreported and that there is a definite desire among radiology staff to broaden and advance the knowledge base and culture regarding the safety aspects of ferromagnetic materials related to MRI investigations.6 A recent study found that only highly ferromagnetic bullets can potentially shift in vivo position within MRI magnetic fields and that most bullets are non-ferromagnetic.5 It was concluded that bullet migration within the static magnetic field is minimal and related to the shape and orientation of the bullet.7 It was further shown that the potential ferromagnetic heating effect does not appear significant.8 This indicates that MRI might not be an absolute contraindication if the bullet (and composition) is known.5 Highly ferromagnetic bullets cause large amounts of MRI susceptibility artefacts, which degrade image quality and thus reduce clinical relevance.5
The Shellock MRI safety list is an invaluable resource if the bullet in question is known. Unfortunately, the current list of bullets is very limited and mostly rated for 1.5 Tesla (T) MRI fields. Clinical 3T and 7T MRI machines are becoming more widely utilised and the increased magnetic fields may elicit different or more dramatic interactions.9
Dual-energy computed tomography (DECT) has in recent years demonstrated its ability to characterise urinary calculi,10 quantify coronary calcification,11 contrast agent uptake measurements,12 ventilation-perfusion assessment with Xenon gas,13 and discriminate intracranial haemorrhage from contrast medium.14
Materials composed of different elemental compositions may produce very similar Hounsfield units (HUs) at a given energy, which significantly complicates the differentiation and classification of such materials. Each voxel’s HU represents a linear attenuation coefficient, which is not unique to any given material but rather a function of the material’s composition, its photon energy interaction and its mass density.15 There are no minimum or maximum limitations to the Hounsfield scale.16
The DECT can be defined as the use of attenuation measurements acquired with different energy spectra, in addition to the known changes in attenuation between those spectra, to differentiate and quantify material composition.15 This was initially explored and described by Godfrey Hounsfield, who stated in 1973, ‘Two pictures are taken of the same slice, one at 100 keV and the other at 140 keV … so that areas of high atomic numbers can be enhanced’.17 Alvarez and Macovski further investigated DECT in 197617 and demonstrated that even with polyenergetic X-ray spectra, one could still separate the measured attenuation coefficients into their contributions from the principal forms of photon interaction with matter, namely photoelectric effect and Compton scatter.18
Forensic medicine studies have found DECT capable of differentiating between commonly found foreign materials,19 and identified a potential relationship between a bullet’s ferromagnetic properties and dual-energy index (DEI) value. The DECT has been proven superior to single-source CT while attempting to differentiate between high-density materials because of acquiring data sets at different X-ray beam energies.20
Bullets are made of a variety of materials. Traditionally, bullet cores contain lead, or an antimony-lead alloy, while bullet jackets are made of copper or gilding metal (copper-zinc alloy). Various materials are commonly used in modern bullets, including bismuth, aluminium, bronze, copper, steel, tin, tungsten, plastics and rubber.
The different high-density metals that constitute bullets might be discernible with industrial CT equipment (> 320 kV), but the radiation doses are not compatible with clinical applications in healthcare.21 Bullet characterisation with clinical CT machines has multiple obstacles pertaining to both artefact generation (e.g., high anatomic number, projectile size, metal interfaces, beam hardening and scatter) and reconstruction algorithms.22 The cupping artefact encountered with metals is a subcategory of beam hardening and scatter, leading to misinterpreting the obtained image as a representation of the bullet jacket and core. The ‘core’ portion of the bullet image is a combination of high attenuation and photon starvation, whereas the ‘jacket’ portion is the outer layer composed of scatter artefacts.23
Recent studies have aimed to combine DECT with the extended HU scale (EHUS) to discriminate between highly ferromagnetic and non-ferromagnetic bullets.24,25 The end goal is to determine MRI safety in patients with retained bullets, especially when close to vital structures. Most of these studies have identified a potential relationship, although the results have been inconsistent.
A commonly identified limiting factor has been CT technology differences. The CT machine manufacturer has been constant, but models and machine service ages differed.20,26,27 These studies demonstrated that highly ferromagnetic and non-ferromagnetic bullet DEI numbers clustered together within respected, although inconsistent, ranges.20,26,27 Interscanner variability has not been evaluated among different DECT machines utilising DEI values.
The purpose of this study was to investigate the theoretical DEI variation between different DECT machines while evaluating the potential ferromagnetic properties within the same sample of ex vivo bullets and metal phantoms.
Materials and methods
Study design, settings and projectiles
This experimental ex vivo study comparatively examined metal phantoms and bullets on two different DECT machines. Eight solid rod metal phantoms (12 mm diameter, 5 cm length) were investigated, representing the wide range of metals used in bullet production. The solid rods were bright bar steel, dark bar steel, tungsten, stainless steel, brass, lead, aluminium and copper. The copper rod was excluded from the study after the MRI testing, as it demonstrated strong magnetism and thus represented an alloy rather than pure copper.
Ten unused bullets of various calibres were examined (Table 1). The main elemental composition of projectiles was derived from physical examination. Specific manufacturers and composition were not available at the time of the investigation. These bullets represented commonly available firearm calibres. The individual bullet composition did not factor into the end study results, as each bullet’s magnetic properties were compared with its DEI value.
TABLE 1: Details of the bullets investigated. |
According to suggestions from previous research, some ballistic projectile categories were excluded from the study because of size, composition, and CT spatial resolution restrictions.28 These were air rifle projectiles, shotgun pellets, paintballs and BB gun pellets.
Empirical ferromagnetic testing for each bullet and metal phantom was performed according to methods described in multiple previous studies concerning the deflection angle.29 The projectiles and metal phantoms were individually suspended on a string in front of a marked non-ferrous vertical board at the portal of an MRI machine (Philips 3T Ingenia, Universitas Academic Hospital’s Clinical Imaging Department), and the deflection angles from vertical were documented. Torque value investigation was not performed, as previous studies have indicated a linear relation between deflection angle and torque propensity.24
Three metal phantoms were non-ferrous, three demonstrated marked ferromagnetic properties, and one had intermediate ferromagnetic properties. The copper metal phantom was found to be a ferromagnetic copper alloy and was excluded from further testing.
Five bullets were non-ferromagnetic, one demonstrated mild ferromagnetic properties, and four were markedly ferromagnetic. Two ferromagnetic bullets had a larger deflection angle (exceeding 90°) compared with the ferromagnetic metal phantoms.
A realistic and reproducible study environment was achieved using a Perspex head phantom (large cylinder 16 cm diameter) while evaluating all metal phantoms and bullets. A central cylindrical cavity within the Perspex phantom enabled a reproducible perpendicular central placement of metal phantoms and bullets within the CT machine gantries.
Methods and measurement
The metal phantoms and bullets were individually scanned on two different single-source DECT machines, which varied in manufacturer, production date, operational time, software and post-processing capabilities:
- General Electric CT Discovery 750 HD 64-slice installed at Universitas Academic Hospital, Clinical Imaging Sciences Department (December 2011). Acquisition and processing were performed on the accompanying workstation (AW Volume share 4). Data collection was carried out on AGFA PACS because of technical limitations on the AW Volume share workstation.
- Siemens Healthcare Somatom Definition AS+ 128-slice installed at Pelonomi Academic Hospital, Clinical Imaging Sciences Department (December 2020). Acquisition and processing were performed on the AW. Data collection was performed on the Syngo.via workstation.
The Siemens Somatom CT scanner allowed the EHUS (−10 240 to +30 710) activation during reconstruction. The GE Discovery required a complete system restart after activating the EHUS function, following which the EHUS was continuously active till manually deactivated.
Temporal sequential scanning of the entire scan volume was performed with tube strength at 100 kV and 140 kV on each metal phantom and projectile to allow for similar machine settings. This approach was used because the dual-energy programme settings did not allow for individual tube setting alterations, nor did the Siemens Somatom allow activation of the EHUS within the dual-energy programmes. Metal phantoms and bullets were scanned with the following parameters26: tube voltage 100 kV and 140 kV, exposure 400 mAs, rotation time 0.5 s, pitch 0.6, slice collimation 2 mm × 64 mm × 0.6 mm, slice acquisition 2 mm × 64 mm × 0.6 mm. Reconstruction was on a sharp tissue kernel with EHUS activated. Slice thickness was 1.5 mm on the Siemens Somatom and 1.25 mm on the GE Discovery. Both were set to 1 mm increments.26
The region of interest (ROI) placement30 was central, without incorporating the edge, and covering more than one pixel. Two sets of readings were obtained per bullet and examination, at least five slices apart. The HU readings were recorded for mean, maximum, minimum and standard deviation (s.d.). Figure 1 illustrates the comparative ROI placement for a metal phantom and 9 mm bullets.
|
FIGURE 1: The comparative region of interest (ROI) placement for a (a) bright bar metal phantom, (b) 9 mm Browning (non-ferrous) bullet, and (c) 9 mm Lugar (ferromagnetic). Tube voltage at 100 kV (left) and 140 kV (right); Siemans Somatom (top) and GE Discovery (bottom). |
|
Two independent senior radiology registrars performed all the readings, one of whom was blinded to the ferromagnetic testing. Metal phantoms and bullets were allocated a specific description used throughout the investigation period. To evaluate intra-reader agreement, a second round of scans and readings were performed on the bullets 2–4 weeks after the initial scans.
Dual-energy index
The DECT data were reconstructed and measured using an EHUS at two different energy levels. These readings were then converted to a DEI value using a standard equation26:
For the low kV, the EHUS reading was at 100 kV and the high kV at 140 kV.
The DEI was calculated for all measured HU readings (mean, maximum, minimum). The mean HU readings and the corresponding DEI values were submitted for statistical analyses and used for further interpretation.
Statistical analysis
Descriptive statistics, namely frequencies and percentages for categorical data and medians and percentiles for numerical data, per type of bullet and metal phantom, were calculated. As a result of the small sample sizes, the bootstrap methodology was not appropriate; therefore, the bullet types were compared by means of non-parametric methods. Inter- and intra-rater reliabilities were calculated and described by means of non-parametric methods for paired data, namely the Signed Rank Test. Intraclass correlations (ICC) using the Shrout-Fleiss formula were calculated. The analysis was performed by the Department of Biostatistics, Faculty of the Health Sciences, University of the Free State.
Ethical considerations
Ethical approval to conduct the study was obtained from the University of the Free State Health Sciences Research Ethics Committee (HSREC) and permission to conduct the study at a state hospital was obtained from the Free State Department of Health (ethics approval no.: UFS-HSD2020/0996/2508-0002). No consent was required in this study.
Results
Intra-reader and inter-reader agreement
Inter-reader agreement for the mean DEI values was not statistically significant. The Siemens Somatom had excellent ICC (p-values: ICC for ROI1 = 0.65:0.90900, ROI2 = 1.00:0.92990, ROI3 = 0.64:0.86502, ROI4 = 0.19:−0.00536) and the GE Discovery had good p-values with poor ICC (p-values: ICC for ROI1 = 0.43:−0.65466, ROI2 = 0.92:0.99177, ROI3 = 0.49:−0.29474, ROI4 = 0.56:−0.69093).31 The poor ICCs were negative and correlated with a wider s.d. on the GE Discovery. The metal phantoms demonstrated a wide p-value range and excellent ICC on the Siemens Somatom (p-values: ICC for ROI1 = 0.94:0.99720, ROI2 = 0.08:0.99856), but poor ICC for the GE Discovery (p-values: ICC for ROI1 = 0.47:0.38143, ROI2 = 0.02:0.78236),31 which had a statistically significant variation on the second ROI readings obtained on the GE Discovery.
Intra-reader agreement was not statistically significant for the bullets’ mean DEI values between the two radiology registrars (R1 and R2) with similar p-value ranges on the Siemens Somatom (R1 ROI1 = 0.71, ROI2 = 0.49; R2 ROI1 = 0.56, ROI2 = 0.32), and GE Discovery (R1 ROI1 = 0.06, ROI2 = 0.63; R2 ROI1 = 0.63, ROI2 = 0.2). The proposed reason for the smaller p-values on the GE Discovery is likely because of the wider s.d. range obtained for the mean ROIs. The metal phantoms demonstrated reduced intra-reader agreement with Investigator 2 (R2) obtaining higher p-values on both Siemens Somatom (R1, p = 0.03; R2, p = 0.92) and GE Discovery (R1, p = 0.22; R2, p = 0.69). Investigator 1 (R1) obtained statistically significant ROI readings on the Siemens Somatom.
Metal phantoms
Metal phantoms (Table 2) were used to evaluate the variable DEI values pertaining to the anatomic number of different metals and demonstrated the identified cupping artefact that these high-density materials create. Metal phantoms consisted of single metal solid rods; thus, no core and jacket combination was often encountered with bullets. The obtained images dramatically demonstrated cupping artefacts, and the outer rim varied because of tube voltage and anatomic number. Cupping artefacts can serve as a visual demonstration of beam hardening and photon starvation.
TABLE 2: Metal phantom average Hounsfield unit and dual-energy index values. |
Figure 2 indicates all the metal phantom mean DEI values obtained at the test sites, differentiating between investigators. Figure 3 and Figure 4 demonstrate the magnetic and non-magnetic metal phantom mean DEI values at each site, differentiating between investigators.
|
FIGURE 2: Overall findings for all metal phantom mean dual-energy index values (distinction is made between investigators and each CT machine). |
|
|
FIGURE 3: Metal phantom mean dual-energy index values obtained at Pelonomi Academic Hospital on Siemens Somatom CT machine (distinction is made between investigators and ferromagnetic properties). |
|
|
FIGURE 4: Metal phantom mean dual-energy index values obtained at Universitas Academic Hospital on GE Discovery CT machine (distinction is made between investigators and ferromagnetic properties). |
|
Siemens Somatom metal phantom findings
Metal phantom DEI values for both investigators were within similar ranges (R1: −0.0569 to 0.0546 with average −0.0070; R2: −0.0558 to 0.0547 with average −0.0060). Further analysis of the DEI values demonstrated that the ferromagnetic metals’ DEI ranges were centrally located within the non-ferromagnetic ranges, which is different from previous studies. The average DEI values for the ferromagnetic and non-ferromagnetic metals differed slightly (R1 ferromagnetic: −0.0488, non-ferromagnetic: −0.0569; R2 ferromagnetic: −0.0487, non-ferromagnetic: −0.0558).
GE Discovery metal phantom findings
Metal phantom DEI values for both investigators were once more within similar ranges, although R2 obtained a narrower range (R1: −0.0737 to 0.0064 with average −0.0299; R2: −0.0679 to 0.0006 with average −0.0240). Further analysis of the DEI values demonstrated that the ferromagnetic metals’ DEI ranges were within the upper half of non-ferromagnetic ranges, which is different from previous studies. The average DEI values for the ferromagnetic and non-ferromagnetic metals differed more pronounced (R1 ferromagnetic: −0.0279, non-ferromagnetic −0.0737; R2 ferromagnetic: −0.0142, non-ferromagnetic: −0.0679). Investigator 1 (R1) had a single significant outlier point (−0.1992), which significantly decreased the average DEI value for the ferromagnetic metals.
Ballistic projectiles
The HU measurements (Table 3) were obtained from the central region of each bullet. The ‘jacket’ readings approached HUmax, even with the EHUS. The measured areas of each bullet and phantom depended on the overall diameter and the thickness of the beam hardening artefacts. Thus, the measured areas around the bullet centre varied between most readings.
TABLE 3: Bullet average Hounsfield unit and dual-energy index values. |
Figure 5 indicates all the bullet mean DEI values obtained at the test sites, differentiating between investigators. Figure 6 and Figure 7 demonstrate the magnetic and non-magnetic bullet mean DEI values at each site, differentiating between investigators.
|
FIGURE 5: Overall findings for all bullet mean dual-energy index values (distinction is made between investigators and each CT machine). |
|
|
FIGURE 6: Bullet mean dual-energy index values obtained at Pelonomi Academic Hospital on Siemens Somatom CT machine (distinction is made between investigators and ferromagnetic properties). |
|
|
FIGURE 7: Bullet mean dual-energy index values obtained at Universitas Academic Hospital on GE Discovery CT machine. Distinction is made between investigators and ferromagnetic properties. |
|
Siemens Somatom bullet findings
The bullet mean DEI values of both investigators had wider ranges compared with the metal phantoms (R1: −0.0635 to 0.178 with average −0.0419; R2: −0.1239 to 0.1589 with average −0.0509). Further analysis of the DEI values demonstrated that the ferromagnetic bullet DEI ranges were within the non-ferromagnetic ranges, which was different from previous studies. The average DEI values for the ferromagnetic and non-ferromagnetic bullets differed more significantly than the metal phantoms (R1 ferromagnetic: −0.0574, non-ferromagnetic: −0.0283; R2 ferromagnetic: −0.0658, non-ferromagnetic: −0.0395). The non-ferromagnetic bullets had more pronounced positive outlier DEI values, which increased the average DEI values.
GE Discovery bullet findings
The bullet mean DEI values of both investigators were once more within similar ranges (R1: −0.0787 to 0.1914 with average −0.0417; R2: −0.0796 to 0.2101 with average −0.0511). Further analysis of the DEI values demonstrated that the ferromagnetic bullet DEI ranges were within the non-ferromagnetic ranges, which was different from previous studies. The average DEI values for the ferromagnetic and non-ferromagnetic bullets differed more pronounced (R1 ferromagnetic: −0.0630, non-ferromagnetic: −0.0239; R2 ferromagnetic: −0.0635, non-ferromagnetic: −0.0358). The non-ferromagnetic bullets had more pronounced positive outlier DEI values, which increased the average DEI values.
Discussion
This study aimed to investigate the proposed interscanner variability described as a limiting factor in previous literature pertaining to DECT as a screening tool for ferromagnetic ballistic projectiles because of previous inconsistent findings.20,26,27,30
Ruder et al.30 demonstrated a considerable HU variability between different CT machines, except for brass, which was found to completely attenuate X-ray beams and thought to obscure the differences observed with other materials. Ruder et al.30 also showed that DECT-capable machines had the lowest HU readings among the researched CT machines, although dual-energy was not utilised. The researchers used 80 kV tube voltage and a fixed tube current time product of 130 mAs. These settings were a practical decision because of the CT machine limitations, as 80 kV was the only tube voltage available on all the machines.30 Paulis et al.23 concluded that higher tube voltages and current time products reduce artefact influences on material differentiation, following which they recommended 150 kV tube voltage and 400 mAs exposure.
The conclusion of Ruder et al.30 holds for single-source HU readings, and the current investigation also had variable individual HU readings among the studied DECT machines. However, the use of DEI values was found to be more consistent between the investigated DECT readings. The DEI values are functions of the HU readings and require different energy levels to calculate, which has the result that the CT technology differences are negated. It can be deduced that the CT technology differences affecting the low kV readings, similarly, influence the high kV readings and that this is annulled during the DEI value calculation.
Winklhofer et al.20 concluded that DECT was more reliable at differentiating materials than single-energy CT. They found that the optimal tube voltage combination was 100 kV/140 kV, although exposure was adjusted to produce constant radiation doses. Diallo et al.26 used the same scan parameters and CT manufacturers, although the models differed (Somatom Definition Flash vs Somatom AS+). Both studies identified a DEI difference between ferromagnetic and non-ferromagnetic bullets, although the relations were opposite. Multiple other research studies19,20,23,26,27,30 have been conducted using Siemens models, which do not account for the proposed inter-manufacturer variability. The reason for the variability among results has been ascribed to individual machine differences.
Our investigation was conducted with two completely different DECT machines, which currently appears to be the only study to utilise other manufacturers. We believe the machines represent a broad enough CT technology variation to represent the proposed limitations previously identified in the tabled articles. Our research advances the literature by evaluating this previously suggested interscanner variability.
The machines had variable DEI ranges for the metal phantoms (Figure 2, Figure 3 and Figure 4), with considerable overlap. The Siemens Somatom DEI averaged more positive (−0.0065) compared with the GE Discovery (−0.027), although the latter had narrower ranges. The ferromagnetic phantom DEI ranges (Siemens Somatom average: −0.0488; GE Discovery average: −0.0211) were within the non-ferromagnetic phantoms’ DEI ranges (Siemens Somatom average: −0.0564; GE Discovery average: −0.0708). This is further complicated by the Siemens ferromagnetic readings occupying the middle ranges and the GE ferromagnetic readings occupying the upper half of the non-ferromagnetic ranges. These findings indicate that the machines have a significant variation, and the individual machines might not be able to clearly distinguish ferromagnetic from non-ferromagnetic metals without individually calibrated reference ranges.
The bullets (Figure 5, Figure 6 and Figure 7) had a much better average DEI value agreement between the two DECT machines, with even the outlier points demonstrating similar values. The bullets showed a much narrower DEI range, with a similar ferromagnetic to non-ferromagnetic relationship. The average ferromagnetic DEI values were significantly lower than the average non-ferromagnetic values, similar to Diallo et al.26 and the opposite to Winklhofer et al.30 The current findings demonstrate a more pronounced overlap of DEI values while comparing ferromagnetic to non-ferromagnetic bullets, whereas previous studies had a more pronounced difference with less overlap.
Despite these variations with previous similar research, the main objective of this investigation was to assess variability between different DECT machines. The average bullet DEI values were comparable between the machines (Siemens Somatom: −0.0464; GE Discovery: −0.0464). Furthermore, even the average ferromagnetic (Siemens Somatom: −0.0616; GE Discovery: −0.0633) and non-ferromagnetic (Siemens Somatom: −0.0339; GE Discovery: −0.0299) bullet DEI values were comparable and within close proximity to each other. The average DEI values give a false impression of a wide difference between ferromagnetic and non-ferromagnetic values, as the ranges largely overlap. Significant outlier values for the non-ferromagnetic bullets increase the average DEI values, however, the outlier DEI values are similar between the DECT scanners.
Limitations
This research investigation had a couple of limitations worth mentioning and may require future study. Small projectiles and fragments were excluded based on CT spatial resolution limitations.20 Core readings were used for DEI value calculation and analysis, as the ‘jackets’ have previously been shown to be unhelpful.20,23 Undamaged unused bullets were used, and optimally placed within the centre of the gantries at perpendicular angles to the X-ray beams. This was done to limit projectile variables and test the research question in isolation. We are aware that bullets will orientate randomly and deform upon impact, and the position related to the gantry and surrounding structures may be different in real life situations. The metal phantom and bullet samples were small; however, we believe that these were sufficiently representative. Larger metal phantom and projectile samples may have increased statistical relevance, although larger studies have been conducted on single machines. Finally, ex vivo analysis is still under investigation and not yet applicable to emergency radiology. Further research and agreement are required before implementation on the living should be considered.
Conclusion
In conclusion, the use of DEI values negates the previous assumption that significant interscanner variability exists among different DECT technologies while assessing highly attenuative ex vivo ballistic projectiles. Previous research indicated that different HU measurements are obtained with different CT machines, but this has not yet been reassessed on DECT machines utilising DEI values. This investigation has demonstrated that even though HU readings may be variable, the implementation of the DEI equation translates this into comparable values with good interscanner agreement.
Future research on DECT implementation for MRI safety prediction related to bullets should use DEI as a reference rather than HU only.
Acknowledgements
The authors would like to acknowledge Ms T. Mulder, medical editor and writer, Faculty of Health Sciences, University of the Free State, for editorial and technical preparation of the manuscript.
Competing interests
The authors declare that they have no financial or personal relationships that may have inappropriately influenced them in writing this article.
Authors’ contributions
F.A.v.d.M. was the primary researcher of this article and E.L. was the supervisor of the study.
Funding information
This research received no specific grant from any funding agency in the public, commercial or not-for-profit sectors.
Data availability
Data supporting the findings of this study are available from the corresponding author, F.A.v.d.M.
Disclaimer
The views and opinions expressed in this article are those of the authors and do not necessarily reflect the official policy or position of any affiliated agency of the authors.
References
- South African Police Service. SAPS crime statistics 2019/2020 [homepage on the Internet]. Pretoria: South African Police Service; 2020 [cited 2021 Feb 01]. Available from: https://www.saps.gov.za/services/crimestats.php
- Reginelli A, Russo A, Maresca D, Martiniello C, Cappabianca S, Brunese L. Imaging assessment of gunshot wounds. Semin Ultrasound CT MR. 2015;36(1):57–67. https://doi.org/10.1053/j.sult.2014.10.005
- Kong VY, Clarke DL. Analysis of 5 years of morbidity and mortality conferences in a metropolitan South African trauma service. S Afr Med J. 2016;106(7):695–698. https://doi.org/10.7196/SAMJ.2016.v106i7.10549
- Joseph C. Characteristics and outcomes of gunshot-acquired spinal cord injury in South Africa. S Afr Med J. 2017;107(6):518–522. https://doi.org/10.7196/SAMJ.2017.v107i6.12296
- Hackenbroch C, Wafa M, Klinger S, Mauer UM. Magnetic resonance imaging in the presence of projectiles and projectile fragments: Artefacts, image quality, rotation and movement. Magn Reson Imaging. 2019;57:143–150. https://doi.org/10.1016/j.mri.2018.11.019
- Kihlberg J, Hansson B, Hall A, Tisell A, Lundberg P. Magnetic resonance imaging incidents are severely underreported: A finding in a multicentre interview survey. Eur Radiol. 2022;32(1):477–488. https://doi.org/10.1007/s00330-021-08160-w
- Dedini RD, Karacozoff AM, Shellock FG, Xu D, McClellan RT, Pekmezci M. MRI issues for ballistic objects: Information obtained at 1.5-, 3- and 7-Tesla. Spine J. 2013;13(7):815–822. https://doi.org/10.1016/j.spinee.2013.02.068
- Fountain AJ, Corey A, Malko JA, Strozier D, Allen JW. Imaging appearance of ballistic wounds predicts bullet composition: Implications for MRI safety [published correction appears in AJR Am J Roentgenol. 2021;217(2):524]. AJR Am J Roentgenol. 2021;216(2):542–551. https://doi.org/10.2214/AJR.20.23648
- Hoff MN, McKinney A 4th, Shellock FG, et al. Safety considerations of 7-T MRI in clinical practice. Radiology. 2019;292(3):509–518. https://doi.org/10.1148/radiol.2019182742
- Graser A, Johnson TR, Bader M, et al. Dual energy CT characterization of urinary calculi: Initial in vitro and clinical experience. Invest Radiol. 2008;43(2):112–119. https://doi.org/10.1097/RLI.0b013e318157a144
- Schwarz F, Nance JW Jr, Ruzsics B, Bastarrika G, Sterzik A, Schoepf UJ. Quantification of coronary artery calcium on the basis of dual-energy coronary CT angiography. Radiology. 2012;264(3):700–707. https://doi.org/10.1148/radiol.12112455
- Feuerlein S, Heye TJ, Bashir MR, Boll DT. Iodine quantification using dual-energy multidetector computed tomography imaging: Phantom study assessing the impact of iterative reconstruction schemes and patient habitus on accuracy. Invest Radiol. 2012;47(11):656–661. https://doi.org/10.1097/RLI.0b013e31826585bb
- Hwang HJ, Seo JB, Lee SM, et al. Assessment of regional Xenon ventilation, perfusion, and ventilation-perfusion mismatch using dual-energy computed tomography in chronic obstructive pulmonary disease patients. Invest Radiol. 2016;51(5):306–315. https://doi.org/10.1097/RLI.0000000000000239
- Gupta R, Phan CM, Leidecker C, et al. Evaluation of dual-energy CT for differentiating intracerebral hemorrhage from iodinated contrast material staining. Radiology. 2010;257(1):205–211. https://doi.org/10.1148/radiol.10091806
- McCollough CH, Leng S, Yu L, Fletcher JG. Dual- and multi-energy CT: Principles, technical approaches, and clinical applications. Radiology. 2015;276(3):637–653. https://doi.org/10.1148/radiol.2015142631
- Ese Z, Kressmann M, Kreutner J, Schaefers G, Erni D, Zylka W. Influence of conventional and extended CT scale range on quantification of Hounsfield units of medical implants and metallic objects. Tech Mess. 2018;85(5):343–350. https://doi.org/10.1515/teme-2017-0122
- Hounsfield GN. Computerized transverse axial scanning (tomography): 1. Description of system. Br J Radiol. 1973;46(552):1016–1022. https://doi.org/10.1259/0007-1285-46-552-1016
- Alvarez RE, Macovski A. Energy-selective reconstructions in X-ray computerized tomography. Phys Med Biol. 1976;21(5):733–744. https://doi.org/10.1088/0031-9155/21/5/002
- Ruder TD, Thali Y, Bolliger SA, et al. Material differentiation in forensic radiology with single-source dual-energy computed tomography. Forensic Sci Med Pathol. 2013;9(2):163–169. https://doi.org/10.1007/s12024-012-9398-y
- Winklhofer S, Stolzmann P, Meier A, et al. Added value of dual-energy computed tomography versus single-energy computed tomography in assessing ferromagnetic properties of ballistic projectiles: Implications for magnetic resonance imaging of gunshot victims. Invest Radiol. 2014;49(6):431–437. https://doi.org/10.1097/RLI.0000000000000032
- Kumar J, Landheer D, Barnes-Warden J, Fenne P, Attridge A, Williams MA. Inconsistency in 9 mm bullets measured with non-destructive X-ray computed tomography. Forensic Sci Int. 2012;214(1–3):48–58. https://doi.org/10.1016/j.forsciint.2011.07.018
- Boas FE, Fleischmann D. CT artifacts: Causes and reduction techniques. Imaging Med. 2012;4(2):229–240. https://doi.org/10.2217/iim.12.13
- Paulis LE, Kroll J, Heijnens L, et al. Is CT bulletproof? On the use of CT for characterization of bullets in forensic radiology. Int J Legal Med. 2019;133(6):1869–1877. https://doi.org/10.1007/s00414-019-02033-0
- Jackowski C, Lussi A, Classens M, et al. Extended CT scale overcomes restoration caused streak artifacts for dental identification in CT – 3D color encoded automatic discrimination of dental restorations. J Comput Assist Tomogr. 2006;30(3):510–513. https://doi.org/10.1097/00004728-200605000-00027
- Woisetschläger M, Lussi A, Persson A, Jackowski C. Fire victim identification by post-mortem dental CT: Radiologic evaluation of restorative materials after exposure to high temperatures. Eur J Radiol. 2011;80(2):432–440. https://doi.org/10.1016/j.ejrad.2010.06.012
- Diallo I, Auffret M, Deloire L, Saccardy C, Aho S, Ben Salem D. Is dual-energy computed tomography helpful to determinate the ferromagnetic property of bullets? J Forensic Radiol Imaging. 2018;15:21–25. https://doi.org/10.1016/j.jofri.2018.10.001
- Gascho D, Zoelch N, Richter H, Buehlmann A, Wyss P, Schaerli S. Identification of bullets based on their metallic components and X-ray attenuation characteristics at different energy levels on CT. AJR Am J Roentgenol. 2019;213(3):W105–W113. https://doi.org/10.2214/AJR.19.21229
- Eggert S, Kubik-Huch RA, Klarhöfer M, et al. Fairly direct hit! Advances in imaging of shotgun projectiles in MRI. Eur Radiol. 2015;25(9):2745–2753. https://doi.org/10.1007/s00330-015-3646-y
- New PF, Rosen BR, Brady TJ, et al. Potential hazards and artifacts of ferromagnetic and nonferromagnetic surgical and dental materials and devices in nuclear magnetic resonance imaging. Radiology. 1983;147(1):139–148. https://doi.org/10.1148/radiology.147.1.6828719
- Ruder TD, Thali Y, Schindera ST, et al. How reliable are Hounsfield-unit measurements in forensic radiology? Forensic Sci Int. 2012;220(1–3):219–223. https://doi.org/10.1016/j.forsciint.2012.03.004
- Koo TK, Li MY. A guideline of selecting and reporting intraclass correlation coefficients for reliability research [published correction appears in J Chiropr Med. 2017;16(4):346]. J Chiropr Med. 2016;15(2):155–163. https://doi.org/10.1016/j.jcm.2016.02.012
|